| P/N |
Description |
Unit Price(USD) |
Standard |
MOQ. |
| L540904 |
Wavelength range: 200-1000 nm, autoscan ;
Slit width: 4 nm ;
Wavelength accuracy :¡À1.0 nm ;
Wavelength setting repeatability :<=0.5 nm ;
Wavelength display :In steps of 0.1 nm ;
Wavelength calibration: Automatic at start ;
Measurement range: 0-125 %T; -0.097-2.70 A; 0-1999 C (0-1999 F) ;
Photometric accuracy: ¡À0.002 Abs (0 to 0.5Abs), ¡À0.004 Abs (0.5 to 1.0 Abs),¡À0.5 %T ;
Photometric repeatability :¡À0.001 Abs (0 to 0.5Abs), ¡À0.002 Abs (0.5 to 1.0 Abs),¡À0.2 %T ;
Stray light: 0.25 %T (at 220,340,420 nm) ;
Baseline flatness: =¡À0.005 Abs ;
Baseline stability: =0.004 Abs / 30min (at 500 nm, when measured at 2 hours after energization) ;
0%T Line noise: =¡À0.1 %T(at 500 nm, when measured at 2 hours after energization) ;
100%T Line noise: =¡À0.3 %T(at 500 nm, when measured at 2 hours after energization) ;
Light source: Automatic changeover linked with wavelength.Changeover point can be set with UV-Solve application software Automatic changeover linked with wavelength.Changeover point is set at 340 nm.
Display :20 X 2 LCD ;
Data output :Digital, RS232 serial ;
Application software :Windows® - based UV-Solve application software
Mini impact printer: Equipped with mini-printer ;
Power consumption :(220V 10% 50Hz 150VA) or (110 V 10% 60Hz 150VA), switchable ;
Standard sample holder: 4-position sample holder ;
Actual dimensions :Spectrophotometer mainframe 456 W X 375 D X 220 H mm (18.0 W x 14.8 D x 8.7 H in) ;
Net weight: 19.0 Kg (42 lbs.) ;
Shipping dimensions :Paperboard container 625 W X 520 D X 348 H mm (24.6 W x 20.5 D x 13.7 H in) ;
Shipping weight :25.0 Kg (55 lbs.) ; |
3017.00 |
ISO9001 |
1 |
| L540905 |
Wavelength range: 200-1000 nm, autoscan ;
Slit width: 4 nm ;
Wavelength accuracy :¡À1.0 nm ;
Wavelength setting repeatability :<=0.5 nm ;
Wavelength display :In steps of 0.1 nm ;
Wavelength calibration: Automatic at start ;
Measurement range: 0-125 %T; -0.097-2.70 A; 0-1999 C (0-1999 F) ;
Photometric accuracy: ¡À0.002 Abs (0 to 0.5Abs), ¡À0.004 Abs (0.5 to 1.0 Abs),¡À0.5 %T ;
Photometric repeatability :¡À0.001 Abs (0 to 0.5Abs), ¡À0.002 Abs (0.5 to 1.0 Abs),¡À0.2 %T ;
Stray light: 0.25 %T (at 220,340,420 nm) ;
Baseline flatness: =¡À0.005 Abs ;
Baseline stability: =0.004 Abs / 30min (at 500 nm, when measured at 2 hours after energization) ;
0%T Line noise: =¡À0.1 %T(at 500 nm, when measured at 2 hours after energization) ;
100%T Line noise: =¡À0.3 %T(at 500 nm, when measured at 2 hours after energization) ;
Light source: Automatic changeover linked with wavelength.Changeover point can be set with UV-Solve application software Automatic changeover linked with wavelength.Changeover point is set at 340 nm.
Display :20 X 2 LCD ;
Data output :Digital, RS232 serial, centronics parallel ;
Application software :Windows® - based UV-Solve application software
Mini impact printer: / ;
Power consumption :(220V 10% 50Hz 150VA) or (110 V 10% 60Hz 150VA), switchable ;
Standard sample holder: 4-position sample holder ;
Actual dimensions :Spectrophotometer mainframe 456 W X 375 D X 220 H mm (18.0 W x 14.8 D x 8.7 H in) ;
Net weight: 18.5 Kg (41 lbs.);
Shipping dimensions :Paperboard container 625 W X 520 D X 348 H mm (24.6 W x 20.5 D x 13.7 H in) ;
Shipping weight :24.5 Kg (54 lbs.); |
2917.00 |
ISO9001 |
1 |
| L540906 |
Wavelength range: 200-1000 nm, autoscan ;
Slit width: 4 nm ;
Wavelength accuracy :¡À1.0 nm ;
Wavelength setting repeatability :<=0.5 nm ;
Wavelength display :In steps of 0.1 nm ;
Wavelength calibration: Automatic at start ;
Measurement range: 0-125 %T; -0.097-2.70 A; 0-1999 C (0-1999 F) ;
Photometric accuracy: ¡À0.002 Abs (0 to 0.5Abs), ¡À0.004 Abs (0.5 to 1.0 Abs),¡À0.5 %T ;
Photometric repeatability :¡À0.001 Abs (0 to 0.5Abs), ¡À0.002 Abs (0.5 to 1.0 Abs),¡À0.2 %T ;
Stray light: 0.25 %T (at 220,340,420 nm) ;
Baseline flatness: =¡À0.005 Abs ;
Baseline stability: =0.004 Abs / 30min (at 500 nm, when measured at 2 hours after energization) ;
0%T Line noise: =¡À0.1 %T(at 500 nm, when measured at 2 hours after energization) ;
100%T Line noise: =¡À0.3 %T(at 500 nm, when measured at 2 hours after energization) ;
Light source: Automatic changeover linked with wavelength.Changeover point is set at 340 nm.
Display :20 X 2 LCD ;
Data output :Digital, RS232 serial;
Mini impact printer: Equipped with mini-printer ;
Power consumption :(220V 10% 50Hz 150VA) or (110 V 10% 60Hz 150VA), switchable ;
Standard sample holder: 4-position sample holder ;
Actual dimensions :Spectrophotometer mainframe 456 W X 375 D X 220 H mm (18.0 W x 14.8 D x 8.7 H in) ;
Net weight: 19.0 Kg (42 lbs.);
Shipping dimensions :Paperboard container 625 W X 520 D X 348 H mm (24.6 W x 20.5 D x 13.7 H in) ;
Shipping weight :24.0 Kg (53 lbs.); |
2083.00 |
ISO9001 |
1 |
| L540907 |
Wavelength range: 200-1000 nm, autoscan ;
Slit width: 4 nm ;
Wavelength accuracy :¡À1.0 nm ;
Wavelength setting repeatability :<=0.5 nm ;
Wavelength display :In steps of 0.1 nm ;
Wavelength calibration: Automatic at start ;
Measurement range: 0-125 %T; -0.097-2.70 A; 0-1999 C (0-1999 F) ;
Photometric accuracy: ¡À0.002 Abs (0 to 0.5Abs), ¡À0.004 Abs (0.5 to 1.0 Abs),¡À0.5 %T ;
Photometric repeatability :¡À0.001 Abs (0 to 0.5Abs), ¡À0.002 Abs (0.5 to 1.0 Abs),¡À0.2 %T ;
Stray light: 0.25 %T (at 220,340,420 nm) ;
Baseline flatness: =¡À0.005 Abs ;
Baseline stability: =0.004 Abs / 30min (at 500 nm, when measured at 2 hours after energization) ;
0%T Line noise: =¡À0.1 %T(at 500 nm, when measured at 2 hours after energization) ;
100%T Line noise: =¡À0.3 %T(at 500 nm, when measured at 2 hours after energization) ;
Light source: Automatic changeover linked with wavelength.Changeover point is set at 340 nm.
Display :20 X 2 LCD ;
Data output :Digital, RS232 serial, centronics parallel;
Power consumption :(220V 10% 50Hz 150VA) or (110 V 10% 60Hz 150VA), switchable ;
Standard sample holder: 4-position sample holder ;
Actual dimensions :Spectrophotometer mainframe 456 W X 375 D X 220 H mm (18.0 W x 14.8 D x 8.7 H in) ;
Net weight: 18.5 Kg (41 lbs.);
Shipping dimensions :Paperboard container 625 W X 520 D X 348 H mm (24.6 W x 20.5 D x 13.7 H in) ;
Shipping weight :23.5 Kg (52 lbs.); |
1933.00 |
ISO9001 |
1 |
| L540908 |
Wavelength range: 200-1000 nm, autoscan;
Slit width: 2 nm ;
Wavelength accuracy :¡À0.5 nm;
Wavelength setting repeatability :<=0.2 nm ;
Wavelength display :In steps of 0.1 nm ;
Wavelength calibration: Automatic at start ;
Measurement range: 0-125 %T; -0.097-2.70 A; 0-1999 C (0-1999 F) ;
Photometric accuracy: ¡À0.002 Abs (0 to 0.5Abs), ¡À0.004 Abs (0.5 to 1.0 Abs),¡À0.5 %T ;
Photometric repeatability ¡À0.001 Abs (0 to 0.5Abs), ¡À0.002 Abs (0.5 to 1.0 Abs),¡À0.2 %T ;
Stray light:0.15 %T (at 220,340,420 nm) ;
Baseline flatness: =¡À0.005 Abs;
Baseline stability: =0.004 Abs / 30min (at 500 nm, when measured at 2 hours after energization) ;
0%T Line noise: =¡À0.1 %T(at 500 nm, when measured at 2 hours after energization) ;
100%T Line noise: =¡À0.3 %T(at 500 nm, when measured at 2 hours after energization) ;
Light source: Automatic changeover linked with wavelength.Changeover point can be set with UV-Solve application software .
Display :20 X 2 LCD ;
Data output :Digital, RS232 serial;
Application software: Windows - based UV-Solve application software ;
Mini impact printer: Equipped with mini-printer;
Power consumption :(220V 10% 50Hz 150VA) or (110 V 10% 60Hz 150VA), switchable;
Standard sample holder: 4-position sample holder;
Actual dimensions :Spectrophotometer mainframe 456 W X 375 D X 220 H mm (18.0 W x 14.8 D x 8.7 H in) ;
Net weight: 19.0 Kg (42 lbs.);
Shipping dimensions :Paperboard container 625 W X 520 D X 348 H mm (24.6 W x 20.5 D x 13.7 H in) ;
Shipping weight :25.0 Kg (55 lbs.); |
3750.00 |
ISO9001 |
1 |
| L540909 |
Wavelength range: 200-1000 nm, autoscan;
Slit width: 2 nm ;
Wavelength accuracy :¡À0.5 nm;
Wavelength setting repeatability :<=0.2 nm ;
Wavelength display :In steps of 0.1 nm ;
Wavelength calibration: Automatic at start ;
Measurement range: 0-125 %T; -0.097-2.70 A; 0-1999 C (0-1999 F) ;
Photometric accuracy: ¡À0.002 Abs (0 to 0.5Abs), ¡À0.004 Abs (0.5 to 1.0 Abs),¡À0.5 %T ;
Photometric repeatability ¡À0.001 Abs (0 to 0.5Abs), ¡À0.002 Abs (0.5 to 1.0 Abs),¡À0.2 %T ;
Stray light:0.15 %T (at 220,340,420 nm) ;
Baseline flatness: =¡À0.005 Abs;
Baseline stability: =0.004 Abs / 30min (at 500 nm, when measured at 2 hours after energization) ;
0%T Line noise: =¡À0.1 %T(at 500 nm, when measured at 2 hours after energization) ;
100%T Line noise: =¡À0.3 %T(at 500 nm, when measured at 2 hours after energization) ;
Light source: Automatic changeover linked with wavelength.Changeover point can be set with UV-Solve application software .
Display :20 X 2 LCD ;
Data output :Digital, RS232 serial, centronics parallel;
Application software: Windows - based UV-Solve application software ;
Mini impact printer: /;
Power consumption :(220V 10% 50Hz 150VA) or (110 V 10% 60Hz 150VA), switchable;
Standard sample holder: 4-position sample holder;
Actual dimensions :Spectrophotometer mainframe 456 W X 375 D X 220 H mm (18.0 W x 14.8 D x 8.7 H in) ;
Net weight: 18.5 Kg (41 lbs.);
Shipping dimensions :Paperboard container 625 W X 520 D X 348 H mm (24.6 W x 20.5 D x 13.7 H in) ;
Shipping weight :24.5 Kg (54 lbs.); |
3617.00 |
ISO9001 |
1 |
| L540910 |
Wavelength range: 200-1000 nm, autoscan;
Slit width: 2 nm ;
Wavelength accuracy :¡À0.5 nm;
Wavelength setting repeatability :<=0.2 nm ;
Wavelength display :In steps of 0.1 nm ;
Wavelength calibration: Automatic at start ;
Measurement range: 0-125 %T; -0.097-2.70 A; 0-1999 C (0-1999 F) ;
Photometric accuracy: ¡À0.002 Abs (0 to 0.5Abs), ¡À0.004 Abs (0.5 to 1.0 Abs),¡À0.5 %T ;
Photometric repeatability ¡À0.001 Abs (0 to 0.5Abs), ¡À0.002 Abs (0.5 to 1.0 Abs),¡À0.2 %T ;
Stray light:0.15 %T (at 220,340,420 nm) ;
Baseline flatness: =¡À0.005 Abs;
Baseline stability: =0.004 Abs / 30min (at 500 nm, when measured at 2 hours after energization) ;
0%T Line noise: =¡À0.1 %T(at 500 nm, when measured at 2 hours after energization) ;
100%T Line noise: =¡À0.3 %T(at 500 nm, when measured at 2 hours after energization) ;
Light source: Automatic changeover linked with wavelength.Changeover point is set at 340 nm.
Display :20 X 2 LCD ;
Data output :Digital, RS232 serial;
Mini impact printer: Equipped with mini-printer;
Power consumption :(220V 10% 50Hz 150VA) or (110 V 10% 60Hz 150VA), switchable;
Standard sample holder: 4-position sample holder;
Actual dimensions :Spectrophotometer mainframe 456 W X 375 D X 220 H mm (18.0 W x 14.8 D x 8.7 H in) ;
Net weight: 19.0 Kg (42 lbs.);
Shipping dimensions :Paperboard container 625 W X 520 D X 348 H mm (24.6 W x 20.5 D x 13.7 H in) ;
Shipping weight :24.0 Kg (53 lbs.); |
2417.00 |
ISO9001 |
1 |
| L540911 |
Wavelength range: 200-1000 nm, autoscan;
Slit width: 2 nm ;
Wavelength accuracy :¡À0.5 nm;
Wavelength setting repeatability :<=0.2 nm ;
Wavelength display :In steps of 0.1 nm ;
Wavelength calibration: Automatic at start ;
Measurement range: 0-125 %T; -0.097-2.70 A; 0-1999 C (0-1999 F) ;
Photometric accuracy: ¡À0.002 Abs (0 to 0.5Abs), ¡À0.004 Abs (0.5 to 1.0 Abs),¡À0.5 %T ;
Photometric repeatability ¡À0.001 Abs (0 to 0.5Abs), ¡À0.002 Abs (0.5 to 1.0 Abs),¡À0.2 %T ;
Stray light:0.15 %T (at 220,340,420 nm) ;
Baseline flatness: =¡À0.005 Abs;
Baseline stability: =0.004 Abs / 30min (at 500 nm, when measured at 2 hours after energization) ;
0%T Line noise: =¡À0.1 %T(at 500 nm, when measured at 2 hours after energization) ;
100%T Line noise: =¡À0.3 %T(at 500 nm, when measured at 2 hours after energization) ;
Light source: Automatic changeover linked with wavelength.Changeover point is set at 340 nm.
Display :20 X 2 LCD ;
Data output :Digital, RS232 serial, centronics parallel;
Power consumption :(220V 10% 50Hz 150VA) or (110 V 10% 60Hz 150VA), switchable;
Standard sample holder: 4-position sample holder;
Actual dimensions :Spectrophotometer mainframe 456 W X 375 D X 220 H mm (18.0 W x 14.8 D x 8.7 H in) ;
Net weight: 18.5 Kg (41 lbs.);
Shipping dimensions :Paperboard container 625 W X 520 D X 348 H mm (24.6 W x 20.5 D x 13.7 H in) ;
Shipping weight :23.5 Kg (52 lbs.); |
2300.00 |
ISO9001 |
1 |